Study on measuring device arrangement of array-type CdTe detector for BNCT-SPECT
Study on measuring device arrangement of array-type CdTe detector for BNCT-SPECT
复制标题
DOI:
10.1016/j.rpor.2015.04.002
复制
发表时间:
2016-03-01
影响因子:
1.2
通讯作者:
Murata, Isao
中科院分区:
文献类型:
--
作者:
Manabe, Masanobu;Nakamura, Soichiro;Murata, Isao
Aim: To design the measuring device arrangement of array-type CdTe detector for BNCT-SPECT.Background: In a boron neutron capture therapy, a very serious unsolved problem exists, namely that the treatment effect for BNCT cannot be known during irradiation in real time. Therefore, we have been developing a so-called BNCT-SPECT with a CdTe detector, which can obtain a three-dimensional image for the BNCT treatment effect by measuring 478 keV gamma-rays emitted from the excited state of 7Li nucleus created by the B-10(n,alpha) reaction. However, no practical uses were realized at present, because BNCT-SPECT requires very severe conditions for spatial resolution, measuring time, statistical accuracy and energy resolution.Materials and methods: The design study was performed with numerical simulations carried out by a 3-dimenaional transport code, MCNP5 considering the detector assembly, irradiation room and even arrangement of arrayed CdTe crystals.Results: The estimated count rate of 478 keV gamma-rays was sufficiently large being more than the target value of over 1000 counts/h. However, the S/N ratio did not meet the target of S/N > 1. We confirmed that deterioration of the S/N ratio was caused by the influence of Compton scattering especially due to capture gamma-rays of hydrogen. Theoretical calculations were thereafter carried out to find out whether anti-Compton measurement in an array-type CdTe detector could decrease the noise due to Compton scatterings.Conclusions: The calculation result showed that the anti-coincidence would possibly increase the S/N ratio. In the next phase, an arrayed detector with two CdTe crystals will be produced to test removal possibility of the anti-coincident event. (C) 2015 Greater Poland Cancer Centre. Published by Elsevier Sp. z o.o. All rights reserved.