Fluorimetric determination of aluminium in serum.
Fluorimetric determination of aluminium in serum.
复制标题
荧光法测定血清中的铝。
DOI:
10.1039/an9891400839
复制
发表时间:
1989
期刊:
影响因子:
--
通讯作者:
T. Kamiki
中科院分区:
文献类型:
--
作者:
Y. Suzuki;S. Imai;T. Kamiki
A convenient fluorimetric method for the routine determination of aluminium in serum has been developed using lumogallion [4-chloro-3-(2,4-dihydroxyphenylazo)-2-hydroxybenzene-1-sulphonic acid]. Losses of aluminium during deproteinisation of the serum were prevented by treatment with a combination of 20 or 30% m/V trichloroacetic acid (TCA) and 5% m/VTCA. Iron(III) was removed by extraction into chloroform with capriquat (methyltrioctylammonium chloride) as an Fe3+-lumogallion-capriquat ternary complex. The interference from Cu2+ was eliminated by using thiosulphate as a masking agent. The detection limit was 3.6 ng ml-1 and the calibration graph was linear up to 1.4 micrograms ml-1 of aluminium. Using the proposed method, the average concentration of aluminium in the serum of healthy subjects was found to be 6.8 ng ml-1, in agreement with values reported in the literature.