A near-infrared calibration method suitable for quantification of broadband data in humans.
A near-infrared calibration method suitable for quantification of broadband data in humans.
复制标题
一种适合人体宽带数据量化的近红外校准方法。
DOI:
10.1016/j.jneumeth.2010.01.037
复制
发表时间:
2010
影响因子:
3
通讯作者:
Dunn,JeffF
中科院分区:
文献类型:
--
作者:
Zhang,Qiong;Srinivasan,Sathyanarayanan;Wu,Ying;Natah,Siraj;Dunn,JeffF
Broadband near-infrared spectroscopy (bNIRS) is a powerful non-invasive technique for the measurement of hemoglobin. bNIRS systems are relatively simple to construct compared with many near-infrared instruments since they operate on the principle of continuous wave. The advantage of the broadband method is the capacity to model the spectra and to use “the second differential method” to quantify deoxyhemoglobin (HHb). An “anoxia pulse” method can be applied to quantify total haemoglobin (tHb) and tissue oxygen saturation (StO2). A disadvantage is that this calibration method is not suitable for application in humans. In this study, we compared the “anoxia pulse” method with “graded hypoxia” method, which can be applied for human studies, to quantify tHb and StO2. The values obtained with the two methods were respectively (tHb=47.8±2.8 and 49.4±7.7μM, mean±S.D., n=8) and (StO2=72.8±3.7% and 73.2±5.7%, mean±S.D., n=8). There was no significant difference (p<0.05) between the two methods, indicating that the graded hypoxia method could be used for quantification of tHb and StO2in human subjects.