GEM-MSTPC : An active-target type detector in low-pressure He/CO2 mixed gas
GEM-MSTPC : An active-target type detector in low-pressure He/CO2 mixed gas
复制标题
GEM-MSTPC:低压He/CO2混合气体中的主动目标型探测器
DOI:
10.1088/1748-0221/7/03/c03036
复制
发表时间:
2012
期刊:
影响因子:
--
通讯作者:
et al
中科院分区:
文献类型:
--
作者:
H.Ishiyama;et al
We developed an active-target type gas counter operating with low pressure He/CO 2 (10%) detector gas for application in studying low-energy nuclear reactions using radioactive nuclear beams (RNBs). A 400-μm-thick gas electron multiplier (THGEM) was used as the proportional counter for high injection rate capability. We examined the gas gain stability and the influence of ion feedback on particle tracks at high beam injection rates of up to 10 5 particles per second (pps) using a low-energy 12 C beam. By means of modifications to the THGEM, the shift in the relative pulse height of the output signal normalized to that measured at the 12 C injection rate of several hundred pps improved to within 2% at the rate of 10 5 pps. The position distortion induced by ion feedback was suppressed to within 1.5 mm because of the relatively high drift field value of 1 kV/cm⋅ atm and the use of the double GEM configuration. The THGEM was found to be applicable for our active target at high injection rates of up to 10 5 pps.