We compared several alternative ECL solutions for Western blot detection of endogenous proteins in whole cell lysates using inexpensive, commercially available reagents. Starting from an existing protocol based on p-coumaric acid (pCA) as enhancer, we found that the ECL solution containing 4-iodophenylboronic acid (41PBA) generated strong specific signals and low background chemiluminescence. We optimised the lummol, 41PBA and hydrogenperoxide concentrations of this 41PBA-ECL solution. The optimised 41PBA-ECL solution (100 mM Tris/HCl pH 8.8, 1.25 mM luminol, 2 mM 41PBA, 5.3 mM hydrogenperoxide) shows a greatly increased signal intensity compared to the initial pCA-ECL protocol and to some commercially available ECL solutions. In addition, the optimised 41PBA-ECL solution also generates much lower background chemiluminescence than other noncommercial ECL solutions using p-coumaric acid or 4-iodophenol as enhancers. The 41PBA-ECL solution was stable when stored but had the lowest background when prepared freshly from stock solutions. Thus, we present an optimised protocol for a well-performing inexpensive ECL solution which is an alternative to expensive commercial ECL solutions and which achieves a better signal and lower background than the commercial solutions tested. (c) 2006 Elsevier B.V. All rights reserved.