Breast tumor imaging using a tiltable head SPECT camera
Breast tumor imaging using a tiltable head SPECT camera
复制标题
DOI:
10.1109/nssmic.2000.949385
复制
发表时间:
2000-10
期刊:
影响因子:
--
通讯作者:
B. C. Pieper;J. Bowsher;M. Tornai;G. Peter;K. Greer;R. Jaszczak
中科院分区:
文献类型:
--
作者:
B. C. Pieper;J. Bowsher;M. Tornai;G. Peter;K. Greer;R. Jaszczak
An axially Tiltable-Head SPECT (TH-SPECT) system with 2 heads was used to image fillable breast and torso phantoms containing multiple lesions at various tilt angles. Breast, liver, and myocardial activity were included in order to simulate direct contamination and Compton scattering expected in clinical scans. High count planar images were acquired for comparison with TH-SPECT, and the TH-SPECT data was reconstructed using an OS-FM algorithm which accounted for the tilted geometry. In order to characterize axial blurring effects inherent with TH-SPECT reconstructions, two cylindrical disk Defrise phantoms, one large Defrise phantom and one mini-Defrise phantom placed inside the fillable breast phantom, were imaged at various tilt angles. Reconstructions of the combined fillable breast and torso phantoms containing two 1.15 ml lesions, one centered axially and one proximal to the anterior chest wall within the breast, were most clearly visible in the 30/spl deg/ reconstructed TH-SPECT images compared to the high-count planar images, providing lesion SNR and contrast improvements of nearly three times.