Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: first validation versus invasive coronary angiography

Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: first validation versus invasive coronary angiography
复制标题

DOI:
10.1007/s00259-011-1877-y
复制
发表时间:
2011-11-01
影响因子:
9.1
通讯作者:
Kaufmann, Philipp A.
Kaufmann, Philipp A.
中科院分区:
医学1区
文献类型:
--
作者:
Fiechter, Michael;Ghadri, Jelena R.;Kaufmann, Philipp A.

文献摘要

被引文献

相似文献

目的评价一种新的混合单光子发射计算机断层扫描(SPECT)/CT装置对衰减校正核心肌灌注成像(MPI)诊断冠状动脉疾病(CAD)的准确性。侵入性冠状动脉造影作为标准的reference.Methods的研究人群包括66例患者(79%的男性,平均年龄63 +/- 11岁)谁接受了1天的Tc-99 m-tetrofosmin药理学负荷/休息检查和血管造影在3个月内。计算了CT X线衰减校正CZT MPI检测CAD的敏感性、特异性、阳性预测值(PPV)和阴性预测值(NPV)以及准确性(每千日元50%管腔狭窄)。敏感性,特异性,PPV,NPV和准确性分别为87,67,92,53和83%.Conclusion在这第一份报告中,CZT SPECT/CT MPI比较与血管造影,我们证实了血管造影记录的CAD检测的准确性高。
Purpose We evaluated the diagnostic accuracy of attenuation corrected nuclear myocardial perfusion imaging (MPI) with a novel hybrid single photon emission computed tomography (SPECT)/CT device consisting of an ultrafast dedicated cardiac gamma camera with cadmium-zinc-telluride (CZT) solid-state semiconductor detectors integrated onto a multislice CT scanner to detect coronary artery disease (CAD). Invasive coronary angiography served as the standard of reference.Methods The study population included 66 patients (79% men; mean age 63 +/- 11 years) who underwent 1-day Tc-99m-tetrofosmin pharmacological stress/rest examination and angiography within 3 months. Sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) as well as accuracy of the CT X-ray based attenuation corrected CZT MPI for detection of CAD (a parts per thousand yen50% luminal narrowing) was calculated on a per-patient basis.Results The prevalence of angiographic CAD in the study population was 82%. Sensitivity, specificity, PPV, NPV and accuracy were 87, 67, 92, 53 and 83%, respectively.Conclusion In this first report on CZT SPECT/CT MPI comparison versus angiography we confirm a high accuracy for detection of angiographically documented CAD.