文献摘要

被引文献

相似文献

我们提出了高Q值射频平面微线圈的设计,制造和测试的核磁共振(NMR)光谱在小体积样品。线圈是用高纵横比SU-8光环氧树脂和电镀铜在玻璃晶片上制作的。晶圆上的电气特性表明,在800 MHz的质量因数高达40。将在300 MHz下测量的品质因数为24的500 μ m直径的微线圈安装在印刷电路板上,并使用铝丝接合进行电接触。将该探针插入7 T超导磁体中,并在单次扫描中获得放置在微线圈上的毛细管中所含的160 nL NMR谱。这项工作是将核磁共振检测集成到微全分析系统(muTAS)的重要一步。(C)2002 Elsevier Science B. V.保留所有权利。
We present the design, fabrication and test of high-Q factor radiofrequency planar microcoils for nuclear magnetic resonance (NMR) spectroscopy in small volume samples. The coils are fabricated on glass wafers using high-aspect ratio SU-8 photoepoxy and copper electroplating. On-wafer electrical characterization shows quality factors up to 40 at 800 MHz. A 500 mum diameter microcoil with a measured quality factor of 24 at 300 MHz is mounted on a printed circuit board and electrically contacted using aluminum wire bonding. This probe is inserted in a 7 T-superconducting magnet, and a H-1-NMR spectrum of 160 nl ethylbenzene contained in a capillary placed over the microcoil is acquired in a single scan. This work is an important step towards the integration of NMR detection into micro-total analysis s' stems (muTAS). (C) 2002 Elsevier Science B.V. All rights reserved.