Measurement of 25-OH-vitamin D in human serum using liquid chromatography tandem-mass spectrometry with comparison to radioimmunoassay and automated immunoassay
Measurement of 25-OH-vitamin D in human serum using liquid chromatography tandem-mass spectrometry with comparison to radioimmunoassay and automated immunoassay
复制标题
DOI:
10.1016/j.jchromb.2010.03.035
复制
发表时间:
2010-05-01
影响因子:
3
通讯作者:
van Daal, Henny
中科院分区:
文献类型:
--
作者:
van den Ouweland, Johannes M. W.;Beijers, Antonius M.;van Daal, Henny
The plasma 25-OH vitamin D concentration is a reliable biomarker for vitamin D status but assay's variability makes adequate monitoring of vitamin D status difficult. We employed isotope-dilution liquid chromatography (LC) tandem-mass spectrometry (MS/MS) for the measurement of both 25-OH vitamin D(3) and 25-OH vitamin D(2) in human serum. Hexadeuterium labelled 25-OH vitamin D(3) internal standard (IS) was added to calibrators (prepared in phosphate-buffered saline with 60 g/L albumin), controls or patient sera and 25-OH vitamin D metabolites were released from vitamin D binding protein by adding sodium hydroxide prior to protein precipitation by acetonitrile/methanol (9:1, v/v). Subsequent off-line solid-phase extraction was followed by chromatographic separation on a C-18 column using a water/methanol/ammonium acetate gradient. Detection was by Atmospheric Pressure Electrospray Ionisation (AP-EI) followed by selected reaction monitoring. We compared the LC-MS/MS assay to the DiaSorin radioimmunoassay (RIA) and a recently re-standardised version of an automated electrochemi-luminescent immunoassay (ECLIA) from Roche Diagnostics. We also analysed external quality control samples from the International Vitamin D External Quality Assessment Scheme (DEQAS) for comparison with other participating laboratories using LC-MS. The method was linear from 5 to at least 550 nmol/L with intra- and interday CV's