Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force-Sensing Catheter
Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force-Sensing Catheter
复制标题
DOI:
10.1111/j.1540-8167.2009.01693.x
复制
发表时间:
2010-07-01
影响因子:
2.7
通讯作者:
Reddy, Vivek Y.
中科院分区:
文献类型:
--
作者:
Thiagalingam, Aravinda;D'Avila, Andre;Reddy, Vivek Y.
Methods and Results: Freshly excised hearts from 11 male pigs were perfused and superfused using fresh, heparinized, oxygenated swine blood in an ex vivo model. One-minute ablations were placed using one of 3 different power control strategies (impedance control-15 target impedance drop, and 20 W or 30 W fixed power) and 3 different contact forces (2 g, 20 g, and 60 g) to give a grid of 9 ablation groups. The force sensing catheter (Tacticath (TM), Endosense SA) was irrigated at 17 mL/min for all of the ablations.Of a total 101 ablations, no thrombus formation was noted but popping was seen in 17 lesions. The lesion depth and incidence of pops was 5.0 +/- 1.3 mm /0%, 5.0 +/- 1.6 mm /10% and 6.7 +/- 2.5 mm /45% for the 15 , 20 W, and 30 W groups (P < 0.01), respectively, and 4.4 +/- 1.8 mm /3%, 5.8 +/- 1.6 mm /17% and 6.6 +/- 2.0 mm /37% for the 2 g, 20 g, and 60 g groups, respectively (P < 0.01). The impedance drop in the first 5 seconds was significantly correlated to catheter contact force: 9.7 +/- 9.9 , 22.3 +/- 11.0 , and 41.7 +/- 22.1 , respectively, for the 2 g, 20 g, and 60 g groups (Pearson's r = 0.65, P < 0.01).Conclusion: Catheter contact force has an important impact on both ablation lesion size and the incidence of pops. (J Cardiovasc Electrophysiol, Vol. pp. 806-811, July 2010).