Performance of a novel 43-cm x 43-cm flat-panel detector with CsI:Tl scintillator
Performance of a novel 43-cm x 43-cm flat-panel detector with CsI:Tl scintillator
复制标题
具有 CsI:Tl 闪烁体的新型 43 厘米 x 43 厘米平板探测器的性能
DOI:
--
复制
发表时间:
2004
期刊:
影响因子:
--
通讯作者:
Yoshihiro Ogawa
中科院分区:
文献类型:
--
作者:
T. Yamazaki;T. Tamura;M. Nokita;S. Okada;Shinsuke Hayashida;Yoshihiro Ogawa
We have developed a novel flat-panel detector with CsI:Tl scintillator. The detector consists of a single piece 43cm x 43cm amorphous silicon thin-film transistor (TFT) array with MIS (metal-insulator-semiconductor) photoelectric converter having a pixel pitch of 160μm coated with a needle-like crystal CsI:Tl scintillator. Signal chain was totally revised from current detector utilizing an innovative sensor technology. The novel detector and current detector were equipped to a digital radiography system allowing a quantitative and comparative study. Results show that the novel detector has a linear response covering the radiographic exposure range. It has a moderate modulation transfer function (MTF) sufficient to the radiography tasks and effective to suppress the aliasing. The detective quantum efficiency (DQE) was almost twice than the current detector. The result of contrast-detail phantom exposed with a 1/2x dose level is equivalent to that of current detector with a 1x dose level. These results show that performance of novel detector is superior to and expected to reduce the patient dose in half than current detector due to higher DQE and innovative sensor technology.