Energy resolution of a silicon detector with the RX64 ASIC designed for X-ray imaging

Energy resolution of a silicon detector with the RX64 ASIC designed for X-ray imaging
复制标题

采用专为 X 射线成像设计的 RX64 ASIC 的硅探测器的能量分辨率

DOI:
10.1016/j.nima.2003.07.020
复制
发表时间:
2003
影响因子:
1.4
通讯作者:
P. Wia̧cek
P. Wia̧cek
中科院分区:
物理与天体物理3区
文献类型:
--
作者:
D. Bollini;A. Rodriguez;W. Dąbrowski;A. Garcia;M. Gambaccini;P. Giubellino;P. Grybos;M. Idzik;A. Marzari;L. Montaño;F. Prino;L. Ramello;M. Sitta;K. Swientek;R. Wheadon;P. Wia̧cek

文献摘要

被引文献

相似文献

从硅微带检测器耦合到RX64 ASIC的结果。该系统能够在数字X射线成像中进行单光子计数,可预见应用于双能量乳腺X射线摄影和血管造影。该探测系统的主要特点是低噪声(阈值可低至104keV)、良好的空间分辨率(100 μm ×300 μm)和良好的计数能力(每通道计数可达100万次,最大计数率约为200kHz)。系统的能量分辨率,获得与几个荧光X射线线,进行了说明。
Results from a silicon microstrip detector coupled to the RX64 ASIC are presented. The system is capable of single photon counting in digital X-ray imaging, with foreseen applications to dual energy mammography and angiography. The main features of the detecting system are low noise (operation with threshold as low as ≈4 keV is possible), good spatial resolution (a pixel of 100 μm ×300 μm when oriented edge-on) and good counting rate capability (up to one million counts per channel with a maximum rate of about 200 kHz per channel). The energy resolution of the system, as obtained with several fluorescence X-ray lines, is described.