Disposable biosensor based on cathodic electrochemiluminescence of tris(2,2-bipyridine)ruthenium(II) for uric acid determination
Disposable biosensor based on cathodic electrochemiluminescence of tris(2,2-bipyridine)ruthenium(II) for uric acid determination
复制标题
DOI:
10.1016/j.aca.2013.01.045
复制
发表时间:
2013-04-03
影响因子:
6.2
通讯作者:
Capitan-Vallvey, L. F.
中科院分区:
文献类型:
--
作者:
Ballesta-Claver, J.;Rodriguez-Gomez, R.;Capitan-Vallvey, L. F.
A new method for uric acid (UA) determination based on the quenching of the cathodic ECL of the tris(2,2-bipyridine)ruthenium(II)-uricase system is described. The biosensor is based on a double-layer design containing first tris(2,2-bipyridine)ruthenium(II) (Ru(bpy)(3)(2+)) electrochemically immobilized on graphite screen-printed cells and uricase in chitosan as a second layer. The uric acid biosensing is based on the ECL quenching produced by uric acid over the cathodic ECL caused by immobilized Ru(bpy)(3)(2+) in the presence of uricase. The use of a -1.1 V pulse for 1 s with a dwelling time of 10 s makes it possible to estimate the initial enzymatic rate, which is used as the analytical signal. The Stern-Volmer type calibration function shows a dynamic range from 1.0 x 10(-5) to 1.0 x 10(-3) M with a limit of detection of 3.1 x 10(-6) M and an accuracy of 13.6% (1.0 x 10(-4) M, n=5) as relative standard deviation. Satisfactory results were obtained for urine samples, creating an affordable alternative for uric acid determination. (C) 2013 Elsevier B.V. All rights reserved.