Development of a compact MRI system for trabecular bone volume fraction measurements
Development of a compact MRI system for trabecular bone volume fraction measurements
复制标题
DOI:
10.1002/mrm.20135
复制
发表时间:
2004-08-01
影响因子:
3.3
通讯作者:
Tsuzaki, T
中科院分区:
文献类型:
--
作者:
Kose, K;Matsuda, Y;Tsuzaki, T
A compact MRI system for measuring trabecular bone volume fraction (TBVF) of the calcaneus was developed with the use of a 0.21 T permanent magnet and portable MRI console. The entire system weighed < 600 kg and was installed in a 2 m x 2 m space. Two cross-sectional spin-echo images of a heel acquired with external reference phantoms (total measurement time = 5 min) were used to quantify the TBVF of the calcaneus. The linearity and reproducibility of the measurements were evaluated by means of proton density-adjusted phantoms. Comparative measurements with quantitative ultrasound (QUS) in groups of healthy female volunteers showed a relatively high positive correlation (R-2 = 0.4539,0.2693) between TBVF and the speed of sound (SOS). These results demonstrate the potential of this new system for measuring bone density.