Evaluation of ultrasound tissue damage based on changes in image echogenicity in canine kidney.
Evaluation of ultrasound tissue damage based on changes in image echogenicity in canine kidney.
复制标题
基于犬肾脏图像回声变化的超声组织损伤评估。
DOI:
10.1109/tuffc.2005.1503997
复制
发表时间:
2005
期刊:
影响因子:
--
通讯作者:
Cain,CharlesA
中科院分区:
文献类型:
--
作者:
Seo,Jongbum;Tran,BinhC;Hall,TimothyL;Fowlkes,JBrian;Abrams,GeraldD;O'Donnell,Matthew;Cain,CharlesA
Sufficiently high intensity ultrasound can create hyperechoic regions in an ultrasound image due to local bubble generation. We explore the link between the temporal extent of these hyperechoic regions and tissue damage caused by ultrasound therapy. The decay rate of increased echogenicity from the focal zone in insonated live exteriorized canine kidney was quantified and correlated to the spatial extent of tissue damage. The decay half-time, t/sub half/, defined as the time for echogenicity enhancement to decay by a factor of 2, was observed in all cases to be greater than 41 s in spatial zones in which extensive histological damage was observed. In cases in which the measured t/sub half/ was less than 11 s, the damage was limited to minor hemorrhage, or it was not detected. These t/sub half/ discrimination boundaries of 41 and 11 s were not statistically different for cases in which contrast agent was used to enhance therapeutic efficiency. This was true even though contrast agent infusion significantly reduced the therapy pulse duration threshold for damage production.