4H-SiC Schottky diode arrays for X-ray detection
4H-SiC Schottky diode arrays for X-ray detection
复制标题
DOI:
10.1016/j.nima.2016.10.002
复制
发表时间:
2016-12-21
影响因子:
1.4
通讯作者:
Barnett, A. M.
中科院分区:
文献类型:
--
作者:
Lioliou, G.;Chan, H. K.;Barnett, A. M.
Five SiC Schottky photodiodes for X-ray detection have been electrically characterized at room temperature. One representative diode was also electrically characterized over the temperature range 20 degrees C to 140 degrees C. The performance at 30 degrees C of all five X-ray detectors, in both current mode and for photon counting X-ray spectroscopy was investigated. The diodes were fabricated in an array form such that they could be operated as either a 2x2 or 1x3 pixel array. Although the devices showed double barrier heights, high ideality factors and higher than expected leakage current at room temperature (12 nA/cm(2) at an internal electric field of 105 kV/cm), they operated as spectroscopic photon counting soft X-ray detectors uncooled at 30 degrees C. The measured energy resolution (FWHM at 17.4 keV, Mo K alpha) varied from 1.36 to 1.68 keV among different diodes.