Photoelectron track image of capillary gas proportional counter
Photoelectron track image of capillary gas proportional counter
复制标题
毛细管气体正比计数器光电子径迹图像
DOI:
10.1016/s0168-9002(03)01056-8
复制
发表时间:
2003
影响因子:
1.4
通讯作者:
N. Ujiie
中科院分区:
文献类型:
--
作者:
H. Sakurai;S. Gunji;F. Tokanai;T. Maeda;N. Saitoh;N. Ujiie
The segmented pixel gas detector has been developed with an optical imaging capillary gas proportional counter (CGPC) to obtain an image of an X-ray photoelectron track. The pixel signal is read out with an image-intensified CCD camera combined with the CGPC through the optical system of mirrors. Using the optical imaging CGPC with a penning gas mixture of Ar+8% CH4+2% TMA at 1atm, we have successfully detected clear images of photoelectron tracks for 22keV X-rays. The detection efficiency of the photoelectron track was approximately 95% for the images. The evaluated range of 5.2mm for 22keV X-ray photoelectron was consistent to the published ranges. The optical imaging CGPC with the high-resolution capability is available for an imaging type of X-ray polarimeter and for the field of medical science.