An adiabatic calorimeter for calibrating capsule type thermometers (CSPRTs) in the range 54 K to 273 K
An adiabatic calorimeter for calibrating capsule type thermometers (CSPRTs) in the range 54 K to 273 K
复制标题
用于校准 54 K 至 273 K 范围内的胶囊式温度计 (CSPRT) 的绝热热量计
DOI:
10.1051/ijmqe/2012015
复制
发表时间:
2012
期刊:
影响因子:
--
通讯作者:
A. E. Matarawy
中科院分区:
文献类型:
--
作者:
M. G. Ahmed;A. E. Matarawy
A new comparison system has been constructed for the calibration of capsule-type standard platinum resistance thermometers (CSPRTs) below 273.16 K at the National Institute of Standards (NIS-Egypt). The system can compare two CSPRTs at once. A gold plated comparison block, in which CSPRTs are mounted for calibration, is made from oxygen-free high-conductivity copper. A description of the system and the comparison block along with the comparison method is given. The standard uncertainty related to the temperature control of the system is estimated to be 0.04 mK. The calibrated values for CSPRTs and obtained using the comparison system are in good agreement with those obtained by the direct realization of the low temperature fixed points of the ITS-90 within the combined standard uncertainty.