Helical CT image noise--analytical results.
Helical CT image noise--analytical results.
复制标题
螺旋CT图像噪声——分析结果。
DOI:
10.1118/1.596950
复制
发表时间:
1993
期刊:
影响因子:
3.8
通讯作者:
M. Vannier
中科院分区:
文献类型:
--
作者:
G. Wang;M. Vannier
Helical CT is an important recent development in x-ray CT. In helical CT, planar projection sets are synthesized from raw projection data via interpolation. Among various interpolation schemes, linear interpolation is usually preferred due to its efficiency and performance. In this paper, image noise variance is derived for typical helical CT linear interpolation techniques, including the full scan (FS), under-scan (US), full scan with interpolation (FI), half-scan (HS), half-scan with interpolation (HI) and half-scan with extrapolation (HE) methods. Image noise deviation ratios of helical CT to conventional 360 degrees reconstruction (CR) are tabulated. These are consistent with previously reported simulation results. The theoretical results provide further understanding of helical CT noise performance. It is shown that helical CT image noise deviation is independent of transaxial position, proportional to the raw projection noise deviation, and not affected by the fan angle (approximately for the HE method). Also, helical CT image noise variance is proportional to the area under the square of the reconstruction filter.