Signal-to-noise ratio enhancement of intermolecular double-quantum coherence MR spectroscopy in inhomogeneous fields with phased array coils on a 3 Tesla whole-body scanner.
Signal-to-noise ratio enhancement of intermolecular double-quantum coherence MR spectroscopy in inhomogeneous fields with phased array coils on a 3 Tesla whole-body scanner.
复制标题
在 3 特斯拉全身扫描仪上使用相控阵线圈增强非均匀场中分子间双量子相干磁共振波谱的信噪比。
DOI:
10.1002/jmri.22434
复制
发表时间:
2011
期刊:
影响因子:
--
通讯作者:
Zhong,Jianhui
中科院分区:
文献类型:
--
作者:
Lin,Yanqin;Chen,Zhong;Zhong,Jianhui
PurposeTo improve signal‐to‐noise ratio (SNR) of intermolecular double‐quantum coherence (iDQC) MRS on a 3 Tesla (T) whole‐body scanner.Materials and MethodsA 32‐channel phased array coil was used to acquire iDQC signal of a MRS phantom in the presence of large field inhomogeneity. The obtained individual spectra from the array elements were combined together in the time domain using a multichannel nonparametric singular value decomposition algorithm. The results were compared quantitatively with those acquired with a circularly polarized (CP) head coil.ResultsThe achieved gain in SNR ranges from 1.63 to 2.10 relative to the CP coil, mainly depending on the relative position between the surface of the phased array coil and the voxel of acquisition.ConclusionSNR enhancement of iDQC MRS in inhomogeneous fields on a 3T whole‐body scanner is feasible with phased array coils. This can facilitate iDQC applications of high‐resolution in vivo spectroscopy in the presence of field inhomogeneity for potential disease diagnosis in humans. J. Magn. Reson. Imaging 2011;33:698–703. © 2011 Wiley‐Liss, Inc.