Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures

Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures
复制标题

DOI:
10.1128/jcm.39.11.4169-4171.2001
复制
发表时间:
2001-11-01
影响因子:
9.4
通讯作者:
Verhaegen, J
Verhaegen, J
中科院分区:
医学2区
文献类型:
--
作者:
Zaman, Z;Roggeman, S;Verhaegen, J

文献摘要

被引文献

相似文献

UF-100流式细胞仪和尿试纸结果进行交叉解释,以预测文化的结果。当细菌大于或等于1,000/穆尔、白色细胞大于或等于20/穆尔或白细胞酯酶阳性时,获得最佳阴性预测值。24例假阴性中有9例具有临床意义。因此,UF-100和尿试纸结果不能准确预测培养结果。
UF-100 flow cytometer and urine strip results were cross-interpreted to predict culture outcomes. The best negative predictive value was obtained with bacteria at greater than or equal to1,000/mul, white blood cells at greater than or equal to 20/mul, or leukocyte esterase positivity. Nine of 24 false negatives were clinically significant. Thus, UF-100 and urine strip results do not accurately predict the outcome of cultures.