Disposable electrochemical immunosensor for carcinoembryonic antigen using ferrocene liposomes and MWCNT screen-printed electrode.
Disposable electrochemical immunosensor for carcinoembryonic antigen using ferrocene liposomes and MWCNT screen-printed electrode.
复制标题
DOI:
10.1016/j.bios.2008.10.006
复制
发表时间:
2009-03
影响因子:
12.6
通讯作者:
S. Viswanathan;C. Rani;Annadurai Vijay Anand;J. Ho
中科院分区:
文献类型:
--
作者:
S. Viswanathan;C. Rani;Annadurai Vijay Anand;J. Ho
Disposable electrochemical immunosensor for the detection of carcinoembryonic antigen (CEA) in saliva and serum was developed. Monoclonal anti-CEA antibodies (αCEA) were covalently immobilized on polyethyleneimine wrapped multiwalled carbon nanotubes screen-printed electrode. A sandwich immunoassay was performed with CEA and αCEA tagged ferrocene carboxylic acid encapsulated liposomes (αCEA-FCL). The square wave voltammetry (SWV) was employed to analyze faradic redox responses of the released ferrocene carboxylic acid from the immunoconjugated liposomes on the electrode surface. The magnitude of the SWV peak current was directly related to the concentration of CEA. The calibration curve for CEA concentration was in the range of 5×10−12to 5×10−7gmL−1with a detection limit of 1×10−12gmL−1(S/N=3). This method provides a high precise and sensitive determination of CEA in human blood serum and saliva samples.