Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems
Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems
复制标题
DOI:
10.1089/dia.2008.0109
复制
发表时间:
2009-02-01
影响因子:
5.4
通讯作者:
Gottlieb, Peter A.
中科院分区:
文献类型:
--
作者:
Garg, Satish K.;Smith, James;Gottlieb, Peter A.
Objective: This study evaluated the accuracy and safety of two continuous glucose monitoring (CGM) systems, the SEVEN (R) (DexCom (TM), San Diego, CA) and the Navigator (R) (Abbott Diabetes Care, Alameda, CA), with the YSI laboratory measurements of blood glucose (blood glucose meter manufactured by YSI, Yellow Springs, OH), when worn concurrently in adults with type 1 diabetes.Design and Methods: Fourteen subjects with type 1 diabetes, 33 +/- 6 (mean +/- SD) years old, were enrolled in this study. All subjects wore both sensors concurrently over three consecutive 5-day CGM sessions (15-day wear). On Days 5, 10, and 15, subjects participated in an 8-h in-clinic session where measurements from the CGM systems were collected and compared with YSI measurements every 15 min. At the end of Day 5 and 10 in-clinic sessions, the sensors were removed, and new sensors were inserted for the following CGM session despite the SEVEN system's recommended use for up to 7 days.Results: The mean absolute relative difference (ARD) for the two CGM devices versus YSI was not different: 16.8% and 16.1% for SEVEN and Navigator, respectively (P = 0.38). In the hypoglycemic region (YSI value