Artificial acrylic finger nails may alter pulse oximetry measurement
Artificial acrylic finger nails may alter pulse oximetry measurement
复制标题
DOI:
10.1016/j.resuscitation.2006.11.018
复制
发表时间:
2007-07-01
期刊:
影响因子:
6.5
通讯作者:
Fiedler, Fritz
中科院分区:
文献类型:
--
作者:
Hinkelbein, Jochen;Koehler, Heike;Fiedler, Fritz
Introduction: Pulse oximetry is the most common technique to monitor oxygen saturation (SpO(2)) during intensive care therapy. However, intermittent co-oximetry is still the "gold standard" (SaO(2)). Besides acrylic nails, numerous other factors have been reported to interfere with pulse oximetry. Data of measurements with artificial finger nails are not sufficiently published.Materials and methods: A prospective clinical-experimentat trial in mechanically ventilated and critically ill patients of an ICU was performed. Patients were randomly assigned to either group S (S: Siemens pulse oximeter) or group P (P: Philips pulse oximeter) prior to the measurements. SPO2 was determined in each patient three times alternately in standard (SPO2) and sideways position at the natural nail ((SPO2)-S-N90). For the reference measurements oxygen saturation was measured by means of a haemoximeter (co-oximetry). Thereafter, SPO2 was obtained at the acrylic finger nail in the same way (SPO2)-S-A and (SPO2)-S-A90). Bias was calculated as AS = (SPO2)-S-N-Sa02 and AS = ASpO(2) - SaO2. Accuracy (mean difference) and precision (standard deviation) were used to determine the measurement discrepancy. P < 0.05 was considered significant.Results: Accuracy and precision without acrylic nails applied were comparable to SaO2 in both groups (n.s.). With acrylic nails applied a bias of AS = -1.1 +/- 3.14% for group S (P = 0.00522) and a bias of AS = +0.8 +/- 13.04% for group P was calculated (n.s.).Conclusion: Acrylic finger nails may impair the measurement of oxygen saturation depending on the pulse oximeter used and may cause significant inaccuracy. Hence, removal of artificial acrylic finger nails may be helpful to assure an accurate and precise measurement with pulse oximetry. (C) 2006 Elsevier Ireland Ltd. All. rights reserved.