Characterization of Traveling Wave Ion Mobility Separations in Structures for Loss less Ion Manipulations
Characterization of Traveling Wave Ion Mobility Separations in Structures for Loss less Ion Manipulations
复制标题
DOI:
10.1021/acs.analchem.5b02481
复制
发表时间:
2015-11-17
影响因子:
7.4
通讯作者:
Smith, Richard D.
中科院分区:
文献类型:
--
作者:
Hamid, Ahmed M.;Ibrahim, Yehia M.;Smith, Richard D.
We report on the development and characterization of a traveling wave (TVV)-based Structures for Loss less Ion Manipulations (TW-SLIM) module for ion mobility separations (IMS). The TVV-SLIM module uses parallel arrays of rf electrodes on two closely spaced surfaces for ion confinement, where the rf electrodes are separated by arrays of short electrodes, and using these TVsTs can be created to drive ion motion. In this initial work, TVVs are created by the dynamic application of dc potentials. The capabilities of the TW-SLIM module for efficient ion confinement, lossless ion transport, and ion mobility separations at different rf and TVV parameters are reported. The TW-SLIM module is shown to transmit a wide mass range of ions (m/z 200-2500) utilizing a confining rf waveform (similar to 1 MHz and,300 Vp.p) and low TW amplitudes (