Optimisation of a lozenge-based sensor for detecting impending blockage of urinary catheters.
Optimisation of a lozenge-based sensor for detecting impending blockage of urinary catheters.
复制标题
DOI:
10.1016/j.bios.2021.113775
复制
发表时间:
2021-11
影响因子:
12.6
通讯作者:
R. Heylen;Max Branson;Lauren Gwynne;B. Patenall;Nina Hauschildt;J. Urie;J. Mercer-Chalmers;
中科院分区:
文献类型:
--
作者:
R. Heylen;Max Branson;Lauren Gwynne;B. Patenall;Nina Hauschildt;J. Urie;J. Mercer-Chalmers;
Catheter-associated urinary tract infections resulting from urease-positive microorganisms are more likely to cause a urinary catheter blockage owing to the urease activity of the microbes. Catheter blockage can be dangerous and increases the risk of severe infections, such as sepsis. Ureases, a virulence factor inProteus mirabilis, cause an increase in urine pH - leading to blockage. An optimised biosensor “lozenge” is presented here, which is able to detect impending catheter blockage. This lozenge has been optimised to allow easy manufacture and commercialisation. It functions as a sensor in a physiologically representative model of a catheterised urinary tract, providing 6.7 h warning prior to catheter blockage. The lozenge is stable in healthy human urine and can be sterilized for clinical use by ethylene oxide. Clinically, the lozenge will provide a visible indication of impending catheter blockage, enabling quicker clinical intervention and thus reducing the morbidity and mortality associated with blockage.