INHERENT LIMITATIONS OF NONDESTRUCTIVE CHLOROPHYLL METERS - A COMPARISON OF 2 TYPES OF METERS
INHERENT LIMITATIONS OF NONDESTRUCTIVE CHLOROPHYLL METERS - A COMPARISON OF 2 TYPES OF METERS
复制标题
DOI:
10.21273/hortsci.27.1.69
复制
发表时间:
1992-01-01
期刊:
影响因子:
1.9
通讯作者:
BUGBEE, B
中科院分区:
文献类型:
--
作者:
MONJE, OA;BUGBEE, B
Two types of nondestructive chlorophyll meters were compared with a standard, destructive chlorophyll measurement technique. The nondestructive chlorophyll meters were 1) a custom built, single-wavelength meter, and 2) the recently introduced, dual-wavelength, chlorophyll meter from Minolta (model SPAD-502). Data from both meters were closely correlated with destructive measurements of chlorophyll (r2 = 0.90 and 0.93; respectively) for leaves with chlorophyll concentrations ranging from 100 to 600 mg.m-2, but both meters consistently overestimated chlorophyll outside this range. Although the dual-wavelength meter was slightly more accurate than the single-wavelength meter (higher r2), the light-scattering properties of leaf cells and the nonhomogeneous distribution of chlorophyll in leaves appear to limit the ability of all meters to estimate in vivo chlorophyll concentration.