Improved coverage in dynamic contrast-enhanced cardiac MRI using interleaved gradient-echo EPI
Improved coverage in dynamic contrast-enhanced cardiac MRI using interleaved gradient-echo EPI
复制标题
DOI:
10.1002/mrm.1910390403
复制
发表时间:
1998-04-01
影响因子:
3.3
通讯作者:
Epstein, FH
中科院分区:
文献类型:
--
作者:
Ding, SJ;Wolff, SD;Epstein, FH
An interleaved gradient-echo echo-planar imaging (IGEPI) sequence was modified for and applied to dynamic contrast-enhanced imaging of the heart. Using IGEPI, images with 3.0 x 3.9 mm nominal in-plane resolution are acquired in 100 ms, enabling eight slices per heartbeat for a heart rate of 60 beats/min. The acquisition speed and use of saturation pre; pulses allows acquisition of short-and long-axis images during the same contrast bolus. IGEPI maintains the acquisition characteristics required for performing a quantitative first-pass perfusion analysis as well as providing improved coverage compared with conventional fast gradient echo.