Accuracy of real-time MR temperature mapping in the brain: A comparison of fast sequences
Accuracy of real-time MR temperature mapping in the brain: A comparison of fast sequences
复制标题
DOI:
10.1016/j.ejmp.2009.11.006
复制
发表时间:
2010-10-01
期刊:
影响因子:
--
通讯作者:
Schick, Fritz
中科院分区:
文献类型:
--
作者:
Kickhefel, A.;Roland, Joerg;Schick, Fritz
Purpose: To compare magnetic resonance (MR) thermometry based on the proton resonance frequency (PRF) method using a single shot echoplanar imaging (ss EPI) sequence to both of the standard sequences, gradient echo (GRE) and segmented echoplanar imaging (seg EPI) in the in vivo human brain, at 1.5T and 3T.Material and methods: Repetitive MR thermometry was performed on the brain of six volunteers using GRE, seg EPI, and ss EPI sequences on whole-body 1.5T and 31 clinical systems using comparable acquisition parameters. Phase stability and temperature data precision in the human head were determined over 12 min for the three sequences at both field strengths. An ex-vivo swine skeletal muscle model was used to evaluate temperature accuracy of the ss EPI sequence during heating by high intensity focused ultrasound (HIFU).Results: In-vivo examinations of brain revealed an average temperature precision of 0.37 degrees C/0.39 degrees C/0.16 degrees C at 3T for the GRE/seg EPI/ss EPI sequences. At 1.5T, a precision of 0.58 degrees C/0.63 degrees C/0.21 degrees C was achieved. In the ex-vivo swine model, a strong correlation of temperature data derived using ss EPI and GRE sequences was found with a temperature deviation