An improved external loop resonator for in vivo L-band EPR spectroscopy
An improved external loop resonator for in vivo L-band EPR spectroscopy
复制标题
DOI:
10.1016/s1090-7807(03)00175-7
复制
发表时间:
2003-09-01
影响因子:
2.2
通讯作者:
Swartz, HM
中科院分区:
文献类型:
--
作者:
Salikhov, I;Hirata, H;Swartz, HM
An improved external loop resonator (ELR) used for L-band electron paramagnetic resonance (EPR) spectroscopy is reported. This improvement is achieved by shortening the parallel coaxial line. The resonant structure is formed by two single turn coils (10 mm in diameter) that are connected to a parallel coaxial line. A resonance frequency of 1197 MHz and a quality factor of 466 were obtained in the absence of biological tissue and similar to1130 MHz and similar to50 with a living animal, respectively. The sensitivity of the new ELR was compared to the previously developed ELR using three types of EPR samples: (1) paramagnetic material with no biological tissue, (2) paramagnetic material in a leg and in the peritoneal cavity of a dead rat, and (3) paramagnetic material in the back of an anesthetized rat. The sensitivity was 1.2-1.6 times greater in the rat and 4.2 times without tissue. (C) 2003 Elsevier Science (USA). All rights reserved.