Improving LIGO calibration accuracy by using time-dependent filters to compensate for temporal variations
Improving LIGO calibration accuracy by using time-dependent filters to compensate for temporal variations
复制标题
通过使用瞬态滤波器补偿时间变化来提高 LIGO 校准精度
DOI:
10.1088/1361-6382/acabf6
复制
发表时间:
2023
影响因子:
3.5
通讯作者:
Wade, L
中科院分区:
文献类型:
--
作者:
Wade, M;Viets, A D;Chmiel, T;Stover, M;Wade, L
The response of the Advanced Laser Interferometer Gravitational-wave Observatory (Advanced LIGO) interferometers is known to vary with time (Tuyenbayev et al 2017 Class. Quantum Grav. 34 015002). Accurate calibration of the interferometers must therefore track and compensate for temporal variations in calibration model parameters. These variations were tracked during the first three Advanced LIGO observing runs, and compensation for some of them has been implemented in the calibration procedure. During the second observing run, multiplicative corrections to the interferometer response were applied while producing calibrated strain data both in real time and in high latency. In a high-latency calibration produced after the second observing run and during the entirety of the third observing run, a correction requiring periodic filter updates was applied to the calibration–the time dependence of the coupled cavity pole frequency