Verification for dose estimation performance of a Monte-Carlo based treatment planning system in University of Tsukuba

Verification for dose estimation performance of a Monte-Carlo based treatment planning system in University of Tsukuba
复制标题

DOI:
10.1016/j.apradiso.2020.109222
复制
发表时间:
2020-12-01
影响因子:
1.6
通讯作者:
Sakae, Takeji
Sakae, Takeji
中科院分区:
工程技术3区
文献类型:
--
作者:
Kumada, Hiroaki;Takada, Kenta;Sakae, Takeji

文献摘要

被引文献

相似文献

筑波大学不仅正在开发用于BNCT的基于直线加速器的中子源(iBNCT),而且还在开发用于BNCT的多模式治疗计划系统(Tsukuba-Plan)。我们目前正在进行多项核查。利用Tsukuba-Plan对iBNCT中的体模实验进行了模拟,并将模拟结果与实验测量结果进行了比较。计算结果与实测值吻合较好。结果表明,Tsukuba-Plan可以正确估计iBNCT的BNCT治疗剂量。
The University of Tsukuba is developing not only a linac-based neutron source for BNCT (iBNCT) but also a multi-modal treatment planning system (Tsukuba-Plan) for BNCT. We are currently performing several verifications. Phantom experiments performed in iBNCT were simulated by the Tsukuba-Plan, and the calculation results were compared with the measurements from the experiments. The calculations were in good agreement with the measurements. The results demonstrated that the Tsukuba-Plan can perform to estimate doses properly for BNCT treatment at iBNCT.