Flight x-ray CCD selection for the AXAF CCD Imaging Spectrometer
Flight x-ray CCD selection for the AXAF CCD Imaging Spectrometer
复制标题
AXAF CCD 成像光谱仪的飞行 X 射线 CCD 选择
DOI:
10.1117/12.255993
复制
发表时间:
1996
期刊:
影响因子:
--
通讯作者:
J. A. Gregory
中科院分区:
文献类型:
--
作者:
M. Pivovaroff;S. Kissel;M. Bautz;G. Prigozhin;T. Isobe;J. W. Woo;J. A. Gregory
As part of our program to select and calibrate flight- quality, x-ray CCD detectors for the AXAF CCD imaging spectrometer, we have developed efficient detector screening methods. Our screening protocol, which measures device-level performance parameters (including noise, dark current and charge transfer efficiency) as well as x-ray spectral resolution in the 0.3 - 6 keV band, allows us quickly to identify which of the greater than 30 flight candidate detectors warrant the expenditure of severely limited time available for calibration. The performance criteria used to rank devices are discussed, and the details of the measurement methods are presented. Summary results of the screening measurements are presented for a large sample of devices, and detailed data on selected devices are described. We find that the performance variations among the sample of flight devices to be relatively small but significant.