Field emission arrays for medical x-ray imaging
Field emission arrays for medical x-ray imaging
复制标题
DOI:
10.1063/1.2186388
复制
发表时间:
2006-03-13
影响因子:
4
通讯作者:
Schwoebel, PR
中科院分区:
文献类型:
--
作者:
Schwoebel, PR
Experiments demonstrate the feasibility of using microfabricated cold field emission cathodes in x-ray sources for medical imaging. Electron beam currents of 25 mA at voltages of 25 kV have been used to produce x-ray source focal spot diameters of 0.2 mm. Analysis of both filtered and unfiltered x-ray spectra from molybdenum anodes show clear characteristic x-ray lines. These x-ray source characteristics are consistent with the requirements for human breast imaging. (c) 2006 American Institute of Physics.