Low Radiation Dose and High Image Quality of 320-Row Coronary Computed Tomography Angiography Using a Small Dose of Contrast Medium and Refined Scan Timing Prediction
Low Radiation Dose and High Image Quality of 320-Row Coronary Computed Tomography Angiography Using a Small Dose of Contrast Medium and Refined Scan Timing Prediction
复制标题
使用小剂量造影剂的 320 排冠状动脉计算机断层扫描血管造影的低辐射剂量和高图像质量以及精细的扫描时序预测
DOI:
10.1097/rct.0000000000000951
复制
发表时间:
2020
影响因子:
1.3
通讯作者:
Asayama Yoshiki
中科院分区:
文献类型:
--
作者:
Shirasaka Takashi;Nagao Michinobu;Yamasaki Yuzo;Kojima Tsukasa;Kondo Masatoshi;Hamasaki Hiroshi;Kamitani Takeshi;Kato Toyoyuki;Asayama Yoshiki
ObjectiveThe aim of the study was to investigate the feasibility of coronary computed tomography (CT) angiography with a low kilovoltage peak scan and a refined scan timing prediction using a small contrast medium (CM) dose.MethodsIn protocol A, 120-kVp scanning and a standard CM dose were used. The scan timing was fixed. In protocol B, 80 kVp and a 60% CM dose were used. The scan timing was determined according to the interval from the CM arrival to the peak time in the ascending aorta. We measured the CT number and recorded the radiation dose.ResultsHigher CT numbers were observed in the left circumflex (proximal, P= 0.0235; middle, P= 0.0007; distal, P< 0.0001) in protocol B compared with protocol A. The radiation dose in protocol B was significantly lower than in protocol A (2.2±0.9 vs 4.3±1.7 mSv).ConclusionsLow-contrast, low–radiation dose, high–image quality coronary CT angiography can be performed with low kilovoltage peak scanning and a refined scan timing prediction.