Intraoperative spectral domain optical coherence tomography for vitreoretinal surgery.
Intraoperative spectral domain optical coherence tomography for vitreoretinal surgery.
复制标题
DOI:
10.1364/ol.35.003315
复制
发表时间:
2010-10-15
期刊:
影响因子:
3.6
通讯作者:
Izatt JA
中科院分区:
文献类型:
--
作者:
Tao YK;Ehlers JP;Toth CA;Izatt JA
We demonstrate in vivo human retinal imaging using an intraoperative microscope-mounted optical coherence tomography system (MMOCT). Our optomechanical design adapts an Oculus Binocular Indirect Ophthalmo Microscope (BIOM3), suspended from a Leica ophthalmic surgical microscope, with spectral domain optical coherence tomography (SD-OCT) scanning and relay optics. The MMOCT enables wide-field noncontact real-time cross-sectional imaging of retinal structure, allowing for SD-OCT augmented intrasurgical microscopy for intraocular visualization. We experimentally quantify the axial and lateral resolution of the MMOCT and demonstrate fundus imaging at a 20 Hz frame rate.