Rapid fat suppression in MRI of the breast with short binomial pulses.
Rapid fat suppression in MRI of the breast with short binomial pulses.
复制标题
使用短二项式脉冲在乳房 MRI 中快速抑制脂肪。
DOI:
10.1002/jmri.20737
复制
发表时间:
2006
期刊:
影响因子:
--
通讯作者:
Morrell,GlenR
中科院分区:
文献类型:
--
作者:
Morrell,GlenR
PurposeTo develop a faster method of fat suppression for use in dynamic contrast enhanced MRI of the breast.Materials and MethodsA method of fast fat suppression is presented using spatially nonselective rapid binomial pulses. In contrast to conventional binomial frequency‐selective pulses, these short pulses are applied without interpulse delay, allowing for very rapid spectrally selective excitation.ResultsEffective water excitation and fat suppression were achieved in breast MRI at 3.0 Tesla with total excitation time as low as 160 μsec, which is several times shorter than the excitation time of currently used fat suppression techniques. Rapid fat suppression comes at the expense of increased specific absorption rate (SAR) and mildly degraded quality of suppression. A flexible tradeoff of short imaging time vs. SAR can be made to optimize imaging speed for fat‐suppressed breast MRI.ConclusionRapid binomial pulses can be used for dynamic contrast enhanced breast MRI with excitation times significantly shorter than currently used fat suppression pulses. Shorter excitation time allows more rapid imaging, allowing greater temporal and spatial resolution for characterization of breast lesions. J. Magn. Reson. Imaging 2006. © 2006 Wiley‐Liss, Inc.