Evaluation of particle size distribution of albuterol sulfate hydrofluoroalkane administered via metered-dose inhaler with and without valved holding chambers
Evaluation of particle size distribution of albuterol sulfate hydrofluoroalkane administered via metered-dose inhaler with and without valved holding chambers
复制标题
DOI:
10.1016/s1081-1206(10)61290-2
复制
发表时间:
2005-01-01
影响因子:
5.9
通讯作者:
Lincourt, W
中科院分区:
文献类型:
--
作者:
Crim, C;Holmes, M;Lincourt, W
Background: Administration of inhaled medications via metered-dose inhaler (MDI) to pediatric patients younger than 4 years usually requires use of a holding chamber or spacer with an attached face mask.Objective: To determine the particle size distribution and overall dose of albuterol from the albuterol sulfate hydrofluoroalkane delivered in conjunction with 2 US-marketed valved holding chambers (VHCs) compared with the dose delivered via MDIs without VHCs.Methods: Cascade impaction methods with high-performance liquid chromatography were used to evaluate the fine particle mass (FPM) of albuterol administered without and with the use of 2 commercially available VHCs. Results: Particle size distributions for the 2 VHCs and the control were similar. The mean FPM values for the 2 VHCs and the control were 32, 28, and 30 mug, respectively. Statistical comparison of the FPM shows a similar profile when differences from the albuterol hydrofluoroalkane without a spacer were evaluated.Conclusions: In vitro results obtained under these test conditions demonstrate that all the FPM values for the VHCs tested were within 15% of the control range, a difference that is unlikely to be clinically meaningful. These results do not warrant a change in the recommended dose of albuterol hydrofluoroalkane administered when using the VHCs tested. The use of an MDI in conjunction with a VHC provides a reasonable therapeutic approach for administration of albuterol hydrofluoroalkane to young children and other patients who have difficulty administering the MDI alone.