High-resolution computed tomography with a compact soft x-ray microscope
High-resolution computed tomography with a compact soft x-ray microscope
复制标题
DOI:
10.1364/oe.17.011057
复制
发表时间:
2009-06-22
期刊:
影响因子:
3.8
通讯作者:
Hertz, Hans M.
中科院分区:
文献类型:
--
作者:
Bertilson, Michael;von Hofsten, Olov;Hertz, Hans M.
Computed tomography based on high-resolution soft x-ray microscopy utilizes the natural contrast for biological specimens provided by the water window (lambda = 2.4 - 4.4 nm) and the high resolving power of zone plate objectives. It is capable of revealing the 3D structure of biological specimens at sub-visible-microscopic resolution. To date, the technique has only been available at synchrotron-based microscopes, which limits the researchers access. In the present paper we demonstrate high-resolution soft x-ray tomography with a laboratory zone-plate-based soft x-ray microscope. The specimen, a diatom mounted on a glass capillary, was reconstructed from a tilt series of 53 images covering 180 using a filtered back projection algorithm. The resolution of the tomogram was estimated to a half period of 140 nm using a differential-phase-residual method. Cryo-fixation, increased source brightness and extended-depth-of-focus objectives are important for pushing the resolution of compact systems for biological samples. (C) 2009 Optical Society of America