Fully waterproof two-axis galvanometer scanner for enhanced wide-field optics optical-resolution photoacoustic microscopy
Fully waterproof two-axis galvanometer scanner for enhanced wide-field optics optical-resolution photoacoustic microscopy
复制标题
DOI:
10.1364/ol.380032
复制
发表时间:
2020-02-15
期刊:
影响因子:
3.6
通讯作者:
Kim, Jeehyun
中科院分区:
文献类型:
--
作者:
Lee, Jaeyul;Han, Sangyeob;Kim, Jeehyun
A large field-of-view and fast scanning of photoacoustic microscopy (PAM) relatively have been difficult to obtain due to the water-drowned structure of the system for the transmission of ultrasonic signals. Researchers have widely studied the achievement of a waterproof scanner for dynamic biological applications with a high-resolution and high signal-to-noise ratio. This Letter reports a novel, to the best of our knowledge, waterproof galvanometer scanner-based PAM system with a successfully attainable 9.0 mm x 14.5 mm scan region, amplitude scan rate of 40 kHz, and spatial resolution of 4.9 mu m. The in vivo characterization of a mouse brain in intact-skull microvascular visualization demonstrated its capability in biomedical imaging and is anticipated to be an effective technique for various preclinical and clinical studies. (C) 2020 Optical Society of America