Comparison of two bioelectrical impedance devices and dual-energy X-ray absorptiometry to evaluate body composition in heart failure
Comparison of two bioelectrical impedance devices and dual-energy X-ray absorptiometry to evaluate body composition in heart failure
复制标题
DOI:
10.1111/jhn.12218
复制
发表时间:
2014-12-01
影响因子:
3.3
通讯作者:
Clausell, N.
中科院分区:
文献类型:
--
作者:
Alves, F. D.;Souza, G. C.;Clausell, N.
Background: The utilisation of bioelectrical impedance analysis (BIA) in heart failure can be affected by many factors and its applicability remains controversial. The present study aimed to verify the adequacy of single-frequency BIA (SF-BIA) and multifrequency BIA (MF-BIA) compared to dual-energy x-ray absorptiometry (DEXA) for evaluating body composition in outpatients with heart failure.Methods: In this cross-sectional study, 55 patients with stable heart failure and left ventricle ejection fraction 0.8; P < 0.001), except for fat mass assessed by SF-BIA, which showed a moderate correlation (r = 0.760; P < 0.001). MF-BIA also showed a better agreement with DEXA by Bland-Altman analysis in all measurements. However, both types of equipment showed wide limits of agreement and a significant relationship between variance and bias (Pitmans's test P > 0.05), except MF-BIA for fat-free mass.Conclusions: Compared with DEXA, MF-BIA showed better accuracy than SF-BIA, although both types of equipment showed wide limits of agreement. The BIA technique should be used with caution, and regression equations might be useful for correcting the observed variations, mainly in extreme values of body composition.