Improving accuracy in DOSY and diffusion measurements using triaxial field gradients
Improving accuracy in DOSY and diffusion measurements using triaxial field gradients
复制标题
DOI:
10.1016/j.jmr.2016.06.011
复制
发表时间:
2016-09-01
影响因子:
2.2
通讯作者:
Morris, Gareth A.
中科院分区:
文献类型:
--
作者:
Kiraly, Peter;Swan, Iain;Morris, Gareth A.
NMR measurements of diffusion in solution, whether primarily quantitative, or, (as in DOSY, Diffusion Ordered Spectroscopy) qualitative, can be particularly demanding. Here we show how the use of appropriate transverse (x, y) pulsed field gradients, orthogonal to the more usual z axis pulsed field gradient applied along the long axis of the sample, can greatly reduce two important sources of systematic error in diffusion experiments. These are the extra signal attenuation caused by sample convection, and gradient-dependent signal phase shifts caused by the magnetic field and field-frequency lock disturbances generated by field gradient pulses. (C) 2016 The Authors. Published by Elsevier Inc.