Dynamic B(0) shimming for multiband imaging using high order spherical harmonic shims.

Dynamic B(0) shimming for multiband imaging using high order spherical harmonic shims.
复制标题

DOI:
10.1002/mrm.28438
复制
发表时间:
2021-01
影响因子:
3.3
通讯作者:
Pan JW
Pan JW
中科院分区:
医学3区
文献类型:
--
作者:
Hetherington HP;Moon CH;Schwerter M;Shah NJ;Pan JW

文献摘要

参考文献

被引文献

相似文献

描述并实现了一种在7T时利用人脑中的球谐填隙实现动态逐层和多频带B0填隙的策略。对于薄轴切片,球谐垫片可以被分成一对空间退化的垫片(z-退化和非z-退化),从而只需要1/2的垫片(非z-退化)来进行单切片优化。然而,当组合时,垫片对可用于同时产生相同的面内对称性,但具有作为其z位置的函数的不同幅度。这实现了等同于通过单片逐片优化实现的多频段填隙。所有数据都是在7T时使用球面谐和垫片获得的,该垫片能够通过两个额外的5阶垫片(1-4+)向上垫片至4阶。动态垫片更新使用具有2ms斜坡和约束优化的10A垫片电源来实现,以最大限度地减少涡流。在8个受试者的组中,我们证明:1)动态更新1-4+阶垫片将全脑B0场的标准差从32.4±2.6和24.9±2赫兹降低到15.1±1.7赫兹;2)当只动态更新非z-简并的垫片(14.3±1.5赫兹)时,或者当使用多频带垫片因子2、MBS=2和所有垫片(14.4±2.0赫兹)时,获得了接近相同的性能。高次球面谐波在静态全局匀场的基础上有了很大的改进,并实现了动态多频段匀场,其性能与动态逐片匀场的性能几乎相同。这减少了回声平面成像中的失真。
To describe and implement a strategy for dynamic slice-by-slice and multi-band B0 shimming using spherical harmonic shims in the human brain at 7T. For thin axial slices, spherical harmonic shims can be divided into pairs of shims (z-degenerate and non-z-degenerate) which are spatially degenerate, such that only ½ of the shims (non-z-degenerate) are required for single slice optimizations. However, when combined, the pairs of shims can be used to simultaneously generate the same in-plane symmetries but with different amplitudes as a function of their z location. This enables multi-band shimming equivalent to that achievable by single slice-by-slice optimization. All data were acquired at 7T using a spherical harmonic shim insert enabling shimming up through 4th order with two additional 5th order shims (1st-4th+). Dynamic shim updating was achieved using a 10A shim power supply with 2ms ramps and constrained optimizations to minimize eddy currents. In groups of 8 subjects we demonstrated that: 1) dynamic updating using 1st-4th+ order shims reduced the standard deviation of the B0 field over the whole brain from 32.4±2.6 and 24.9±2Hz with 1st-2nd and 1st-4th+ static global shimming to 15.1±1.7Hz; 2) near equivalent performance was achieved when dynamically updating only the non-z-degenerate shims (14.3±1.5Hz), or when a using multi-band shim factor of 2, MBs=2, and all shims (14.4±2.0Hz). High order spherical harmonics provide substantial improvements over static global shimming and enable dynamic multi-band shimming with near equivalent performance to that of dynamic slice-by-slice shimming. This reduces distortion in echo-planar imaging.
DOI: 10.1002/mrm.25695
发表时间: 2016-03-01
影响因子: 3.3
作者:
Fillmer, Ariane;Vannesjo, Signe Johanna;Henning, Anke
通讯作者: Henning, Anke
DOI: 10.1002/mrm.1910290613
发表时间: 1993-06-01
影响因子: 3.3
作者:
GRUETTER, R
通讯作者: GRUETTER, R
DOI: 10.1002/mrm.27899
发表时间: 2019-08-01
影响因子: 3.3
作者:
Liao, Congyu;Stockmann, Jason;Setsompop, Kawin
通讯作者: Setsompop, Kawin
DOI: 10.1002/mrm.1910360125
发表时间: 1996-07-01
影响因子: 3.3
作者:
Blamire, AM;Rothman, DL;Nixon, T
通讯作者: Nixon, T
DOI: 10.1016/j.neuroimage.2017.06.013
发表时间: 2018-03
期刊: NeuroImage
影响因子: 5.7
作者:
Stockmann JP;Wald LL
通讯作者: Wald LL