Simultaneously determination of bisphenol A and uric acid by zinc/aluminum-layered double hydroxide-2-(2,4- dichlorophenoxy) propionate paste electrode
Simultaneously determination of bisphenol A and uric acid by zinc/aluminum-layered double hydroxide-2-(2,4- dichlorophenoxy) propionate paste electrode
复制标题
DOI:
10.20964/2019.08.75
复制
发表时间:
2019-08
影响因子:
1.5
通讯作者:
Nurul Syahida Mat Rais
中科院分区:
文献类型:
--
作者:
Nurul Syahida Mat Rais
A novel zinc/aluminium-layered double hydroxide-2-(2,4-dichlorophenoxy) propionate (Zn/Al-LDH-DPPA) nanocomposite to modify with multi-walled carbon nanotube (MWCNT) paste electrode for simultaneously determination of uric acid and bisphenol A. The Zn/Al-LDH-DPPA/MWCNT morphology was carried out by a scanning electron microscope and transmission electron microscope, electrode-electrolyte interface properties were examined by electrochemical impedance spectroscopy, and electrochemical responses thoroughly investigated by square-wave voltammetry. The electrode established linear plot for uric acid and bisphenol A from 5.0 x 10 —6 M to 7.0 x 10 —4 M (R 2 = 0.9980) and (R 2 = 0.9960) and the detection limits were 7.95 x 10 —7 M and 8.71 x 10 —7 M (S/N=3). The as-prepared sensor is reproducible, highly stable, sensitive as well as ascribed good anti-interference ability. The Zn/Al-LDH-DPPA/MWCNT electrode is applicable to evaluate simultaneously determination of UA and BPA in lake, river and, urine samples with good percentage recovery.