Automated motion artifact correction for dynamic contrast-enhanced fluorescence imaging during open orthopedic surgery.
Automated motion artifact correction for dynamic contrast-enhanced fluorescence imaging during open orthopedic surgery.
复制标题
开放式骨科手术期间动态对比增强荧光成像的自动运动伪影校正。
DOI:
10.1117/12.2650028
复制
发表时间:
2023
期刊:
影响因子:
--
通讯作者:
Jiang,Shudong
中科院分区:
文献类型:
--
作者:
Tang,Yue;Gitajn,ILeah;Cao,Xu;Han,Xinyue;Elliott,JonathanT;Yu,Xiaohan;Bateman,LoganM;Malskis,BethanyS;Fisher,LillianA;Sin,JessicaM;Henderson,EricR;Pogue,BrianW;Jiang,Shudong
Indocyanine green (ICG)-based dynamic contrast-enhanced fluorescence imaging (DCE-FI) can objectively assess bone perfusion intraoperatively. However, it is susceptible to motion artifacts due to patient’s involuntary respiration during the 4.5-minute DCE-FI data acquisition. An automated motion correction approach based on mutual information (MI) frameby-frame was developed to overcome this problem. In this approach, MIs were calculated between the reference and the adjacent frame translated and the maximal MI corresponded to the optimal translation. The images obtained from eighteen amputation cases were utilized to validate the approach and the results show that this correction can significantly reduce the motion artifacts and can improve the accuracy of bone perfusion assessment.