A Dedicated 36-Channel Receive Array for Fetal MRI at 3T.
A Dedicated 36-Channel Receive Array for Fetal MRI at 3T.
复制标题
用于 3T 胎儿 MRI 的专用 36 通道接收阵列
DOI:
10.1109/tmi.2018.2839191
复制
发表时间:
2018-10
影响因子:
10.6
通讯作者:
Zheng H
中科院分区:
文献类型:
--
作者:
Chen Q;Xie G;Luo C;Yang X;Zhu J;Lee J;Su S;Liang D;Zhang X;Liu X;Li Y;Zheng H
Due to a lack of fetal imaging coils, the standard commercial abdominal coil is often used for fetal imaging, the performance of which is limited by its insufficient coverage, element number, and Signal-to-noise ratio (SNR). In this paper, a dedicated 36-channel coil array, of which size can best fit the body sizes of pregnancy gestation from 20 to 37+ weeks, was designed for fetal imaging at 3T. SNR with full phase encoding and G-factor denoted as noise amplification for parallel imaging were quantitatively evaluated by phantom studies. Compared with a commercial abdominal coil array, the proposed 36-channel fetal array provides not only SNR improvements in full phase encoding (with 10% in the region where the whole fetal body was located, and up to 40% in the edge region where the fetal brain and heart may appear) but also an augmented parallel imaging capability and remarkable SNR improvements at high acceleration factors.