The MR Cap: A single-sided MRI system designed for potential point-of-care limited field-of-view brain imaging
The MR Cap: A single-sided MRI system designed for potential point-of-care limited field-of-view brain imaging
复制标题
DOI:
10.1002/mrm.27861
复制
发表时间:
2019-11-01
影响因子:
3.3
通讯作者:
Wald, Lawrence L.
中科院分区:
文献类型:
--
作者:
McDaniel, Patrick C.;Cooley, Clarissa Zimmerman;Wald, Lawrence L.
Purpose: The size, cost, and siting requirements of conventional MRI systems limit their availability and preclude usage as monitoring or point-of-care devices. To address this, we developed a lightweight MRI for point-of-care brain imaging over a reduced field of view (FOV).Methods: The B-0 magnet was designed with a genetic algorithm optimizing homogeneity over a 3 x 8 x 8 cm FOV and a built-in gradient for slice selection or readout encoding. An external pair of gradient coils enables phase encoding in the other two directions and a radiofrequency (RF) coil provides excitation and detection. The system was demonstrated with high-resolution 1D "depth profiling" and 3D phantom imaging.Results: The lightweight B0 magnet achieved a 64-mT average field over the imaging region at a materials cost of