Diamagnetic Levitation Thermometer
Diamagnetic Levitation Thermometer
复制标题
抗磁悬浮温度计
DOI:
10.1002/tee.23115
复制
发表时间:
2020
影响因子:
1
通讯作者:
Ono Takahito
中科院分区:
文献类型:
--
作者:
Inomata Naoki;Inomata Eiji;Ono Takahito
This paper reports a diamagnetic levitation thermometer using a pyrolytic graphite (PG). The levitation height of diamagnetic materials depends on the temperature. The system is composed of a PG disc, permanent magnet for driving, non‐contact laser displacement meter, and spot heater. We evaluated the height in air and vacuum changing the temperature of the PG disc, and the disc changes its height as to be −1.49 and 1.55 µm/°C in air and vacuum, respectively. Differential thermal analysis of polystyrene is demonstrated. The potential of the diamagnetic levitation as a new principle measuring temperature changes is indicated. © 2020 Institute of Electrical Engineers of Japan. Published by John Wiley & Sons, Inc.