Key technologies of a focused ion beam system for single ion implantation
Key technologies of a focused ion beam system for single ion implantation
复制标题
单离子注入聚焦离子束系统关键技术
DOI:
10.1116/1.590194
复制
发表时间:
1998
影响因子:
1.4
通讯作者:
I. Ohdomari
中科院分区:
文献类型:
--
作者:
T. Matsukawa;T. Shinada;T. Fukai;I. Ohdomari
A focused ion beam (FIB) system was remodeled specially to realize single ion implantation (SII) by which we intended to implant an accurate number of ions one by one into ultrafine semiconductor regions. In the SII, single ions are extracted by chopping the ion beam, and one-by-one extraction of ions have become possible by installing an ultrahigh speed amplifier for chopping. In order to achieve accurate detection of each single ion incidence in the SII, detection of SEs emitted upon ion incidence with a high sensitivity and a high signal to noise (S/N) ratio is essential. Signals from the SE detector which synchronize to an instance of chopping are selectively counted to achieve high S/N ratio. Contaminant particles originating from neutralization and scattering of the ion beam are eliminated by sliding the ion source off the beam axis and cutting off the ion beam at the entrance of the FIB’s mass separator.