Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension
Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension
复制标题
DOI:
10.1089/dia.2008.0032
复制
发表时间:
2009-02-01
影响因子:
5.4
通讯作者:
Chase, H. Peter
中科院分区:
文献类型:
--
作者:
Buckingham, Bruce;Cobry, Erin;Chase, H. Peter
Background: Nocturnal hypoglycemia is a significant problem. From 50% to 75% of hypoglycemia seizures occur at night. Despite the development of real-time glucose sensors (real-time continuous glucose monitor [CGM]) with hypoglycemic alarms, many patients sleep through these alarms. The goal of this pilot study was to assess the feasibility using a real-time CGM to discontinue insulin pump therapy when hypoglycemia was predicted.Methods: Twenty-two subjects with type 1 diabetes had two daytime admissions to a clinical research center. On the first admission their basal insulin was increased until their blood glucose level was